Column Packing is one of the most important step in the chromatographic purification process. A well packed column yields higher resolution and yield.
Column Packing Efficiency can be checked by calculating:
- HETP - Height Equivalent Theoretical Plates
- Peak Asymmetry Factor (AF / As)
HETP Calculation:
HETP = ( d2 * L ) / (5.54* t2)
Where: t = retention value expressed as units of time, volume or measured distance on the Chromatogram
d = peak width at half height expressed in same units as “t”
L = column length (in units of cm for HETP or m for N/m)
Where: t = retention value expressed as units of time, volume or measured distance on the Chromatogram
d = peak width at half height expressed in same units as “t”
L = column length (in units of cm for HETP or m for N/m)
N/m = (5.54* t2) / ( d2 * L )
Asymmetry Factor (AF / As) Calculation:
AF = b/a
Where: a is the peak width (left half) at 10% of peak height, b is the peak width (right half) at 10% of peak height,
Asymmetry Factor (AF / As) Calculation:
AF = b/a
Where: a is the peak width (left half) at 10% of peak height, b is the peak width (right half) at 10% of peak height,
Check out the Column Packing Efficiency Protocol
Column Packing Guidelines - TOSOH Biosciences
References
TOSOH Biosciences - Technical Resources
GE Lifescience - Technical Resources
Pall Lifesciences - Technical Resources